Technical data

Liquids
Measuring PrincipleElectromagnetic
Product headlineThe flowmeter for smallest flow rates with flexible system integration. For demanding hygienic applications.
Sensor featuresFlexible installation concept – numerous hygienic process connections. Energy‐saving flow measurement – no pressure loss due to cross‐section constriction. Maintenance‐free – no moving parts. Liner made of PFA. Sensor housing made of stainless steel (3-A, EHEDG).
Transmitter featuresQuality – software for filling & dosing, density, electrode cleaning and also advanced diagnostics. Easy calculation – bidirectional totalizers. Automatic recovery of data for servicing. 4‐line backlit display with touch control. Device in compact or remote version.
Nominal diameter range DN 2...150
1/12"...6"
Wetted materials Liner: PFA
Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum
Process Connections: stainless steel, 1.4404 (F316L); PVDF; PVC adhesive sleeve
Seals: O-ring seal (EPDM, FKM, Kalrez), aseptic molded seal (EPDM, FKM, silicone)
Grounding Rings: stainless steel, 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); tantalum
Measured variables Volume flow
Max. measurement error±0.2%
Measuring range0...600 m3/h
Max. process pressurePN16 to 40
Medium temperature range-20...+150°C
(-4...+302°F)
Ambient temperature range-20...+60°C
-40...+60°C (optional)
Degree of protection IP 67 (NEMA 4x)
Display/Operation Four line display with backlit
Touch control
Outputs 4...20 mA
Pulse / frequency(active/passive)
Relais
Status
Inputs Status
Current input
Digital communication HART
PROFIBUS PA
Profibus DP
FOUNDATION Fieldbus
Hazardous area approvals ATEX, FM, CSA
Product safety CE, C-tick, EAC marking
Hygienic approvals and certificates 3-A, EHEDG