Technical data

Temperature transmitters
Measuring PrincipleResistance Temperature Detector
Input2 x RTD, TC, Ohm, mV
Output4...20 mA HART
FOUNDATION Fieldbus H1
PROFIBUS PA
Auxiliary power supplyHART:
10,5...40 V DC
Ex ia: 10,5...30 V DC
FOUNDATION Fieldbus:
9...32 V DC
Ex ia: 10,5...24 V DC
PROFIBUS PA:
9...32 V DC
Ex ia: 10,5...24 V DC
CommunicationHART-protocol
FOUNDATION Fieldbus
PROFIBUS PA
Installationin field
Accuracy(Pt100, -50...200 °C) <= 0,15 K (HART)
(Pt100, -58...392 °F) <= 0,3 °F (HART)
(Pt100) <= 0,1 K (FF, PA)
(Pt100) <= 0,18 °F (FF, PA)
Galvanic isolationyes
CertificationUL rec. Comp (HART)
FM IS, NI I/1+2/ABCD
CSA IS, NI I/1+2/ABCD
ATEX II1G Ex ia IIC T6
ATEX II2G Ex d IIC T6
ATEX II3G Ex nA IIC T6
ATEX II3G Ex ic IIC T6
ATEX IECEx II2D Ex tb IIIC Db
ATEX II1G Ex ia IIC T6, II2D Ex ia IIIC
FM XP, NI, DIP I,II,III/1+2/A-G
CSA XP, NI, DIP I,II,III/1+2/A-G
FM XP, DIP, IS, NI I,II,III/1+2/A-G
CSA XP, DIP, IS, NI I,II,III/1+2/A-G
FM+CSA XP,DIP, IS, NI I,II,III/1+2/A-G
CSA General Purpose
IECEx Ex d IIC T6 Gb
IECEx Ex ia, Ex d
NEPSI Ex ia IIC T4-T6
NEPSI Ex nA II T4-T6
NEPSI Ex d IIC T4-T6
TIIS Ex ia IIC T4
TIIS Ex ia IIC T6
TIIS Ex d IIC T6
marine approval only for HART-protocol
SIL2 only for HART-protocol
EAC Ex ia IIC T6 Ga
EAC Ex d IIC T6 Gb